On-stream determination of xanthate concentrations can be made by photometric measurements at 313 nm.A simple double-beam instrument was developed for this purpose. A mercury lamp is used as the light source and a composite filter of UGll glass and chromate solution is used to isolate the required spectral region. The light beam is split into two components of approximately equal intensity. One component is passed through a reference cell containing water to a photo detector; the other component is passed through the xanthate solution to a second photo detector. The reference signal controls amplifier gain to eliminate effects of lamp intensity changes and the amplified sample signal can then be used to measure xanthate concentration.Plant trials showed that provided a clear solution free of air bubbles was obtained from an on-stream filter, the instrument indicated the xanthate concentration and operated satisfactorily for long periods.